Development of a standardized flow cytometric method to conduct longitudinal analyses of intracellular CD3ζ expression in patients with head and neck cancer

  • Authors:
    • Deepak Upreti
    • Alok Pathak
    • Sam K. P. Kung
  • View Affiliations

  • Published online on: February 9, 2016     https://doi.org/10.3892/ol.2016.4209
  • Pages: 2199-2206
Metrics: Total Views: 0 (Spandidos Publications: | PMC Statistics: )
Total PDF Downloads: 0 (Spandidos Publications: | PMC Statistics: )


Abstract

Head and neck squamous cell carcinoma (HNSCC) is the sixth most common neoplasm in the world. The follow‑up protocols currently available do not appear to diagnose treatment failures and recurrences early enough to provide the best treatment to improve the survival rates of patients. The identification of biomarkers may aid in diagnosing, monitoring the progression, or predicting treatment outcomes in HNSCC. The present study aimed to evaluate whether cluster of differentiation (CD) 3ζ chain expression may serve as a biomarker for the early detection of recurrent or persistent HNSCC. However, in a longitudinal study, a standardized method that allows consistent data comparisons in an inter‑assay manner is critical. The present study reveals a method to monitor expression levels of CD3ζ over multiple time‑points using flow cytometry. The present study validated the use of an internal control and normalization procedure for tracking alterations in CD3ζ expression in samples from patients with HNSCC, which were collected and assayed for a longitudinal study.

Introduction

Head and neck squamous cell carcinoma (HNSCC) is located in the upper aerodigestive tract. Although it is the sixth most common neoplasm in the world, advances in multi-modality treatments, including surgery, radiation and chemotherapy, have not increased the 5-year survival rate of ~50% over the past 35 years (1,2). Identifying biomarkers in the blood or other body fluids of an individual may aid in the diagnosis of disease and predict treatment outcomes in patients with HNSCC (3,4). In a cross-sectional study, analysis of a candidate biomarker uses single time-point sampling from population data (5). By contrast, a longitudinal analysis involves using multiple samples from a single patient at various time points; therefore, this type of analysis reveals quantitative alterations in the candidate biomarker in the patient, which may be associated with improved health, treatment outcomes and clinical status of the patient (6). However, to quantify a candidate biomarker in a longitudinal study, a standardized assay that supports robust and consistent inter-assay data collection is critical for statistical analyses (7).

Flow cytometry is a laser-based technology for the analysis of physical and chemical properties of cells or cellular components. Advances in flow cytometer designs and novel fluorochromes for the chemical conjugation of antibodies have facilitated robust, multi-parametric analysis at a single cell level. The technology has been widely used in basic research (8), clinical analysis of patient immune responses (9), hematological malignancies and minimal residue disease (1012), and enumeration of lymphocyte populations in immunodeficient diseases and cancer (13,14).

The present study aimed to evaluate whether the expression of cluster of differentiation (CD) 3ζ in peripheral T cells from patients with HNSCC may serve as a biomarker for the early detection of recurrent or persistent HNSCC in a longitudinal study. The present study used commercially available anti-CD3ζ antibodies in flow cytometry to determine CD3ζ expression in a cohort population of patients with HNSCC prior to and following clinical treatments. However, while comparing the longitudinal data obtained over multiple time points, the present study observed that numerous ranges of CD3ζ expression were exhibited in the flow cytometric assays. The present study developed and described a method that conducts longitudinal flow cytometric analysis of a candidate biomarker, such as CD3ζ, in patients with HNSCC.

Materials and methods

Cohort of patients with HNSCC

In total, 24 patients with HNSCC, which consisted of 15 male and 9 female patients, were evaluated in the current study. The age range of the patients was 21–94 years, with a median age of 64 years. Histological analysis revealed that 95.83% of the patients possessed cancers of a squamous origin. In total, 16 patients were treated using a single treatment modality, including surgery, radiation or chemotherapy, and 8 patients were treated using a combination of various treatment modalities, including surgery with radiation or chemotherapy (Table I). Peripheral blood (10–15 ml) was collected from the 24 patients and also from 24 healthy control donors. Peripheral blood mononuclear cells (PBMCs) were isolated using Ficoll gradient centrifugation, as previously described (15,16). The present study was approved by the Research Ethics Board at the University of Manitoba (Winnipeg, MB, Canada).

Table I.

Clinical pathological characteristics of 24 patients with head and neck squamous cell carcinoma.

Table I.

Clinical pathological characteristics of 24 patients with head and neck squamous cell carcinoma.

Characteristicsn
Total24
Gender
  Male15
  Female  9
Age, years
  Median64
  Range21–94
Histological type
  Squamous cell23
  Merkel cell  1
Treatment
  Surgery13
  Chemo  1
  Surgery + chemo + radiation  4
  Surgery + radiation  1
  Radiation  1
  Chemo + radiation  4

[i] Chemo, chemotherapy.

Antibodies and flow cytometry

Sequential staining was performed to evaluate CD3ζ expression in T cells using flow cytometry. Surface CD3ε, a marker of T cells, was stained using specific fluorescein isothiocyanate (FITC) mouse anti-human CD3 monoclonal antibody (clone, UCHT1; cat no. 55916; dilution, 1:66; BD Biosciences, Franklin Lakes, NJ, USA). Following surface staining, PBMCs were fixed in 1% paraformaldehyde (Sigma-Aldrich, Ontario, Canada)for 10 min at room temperature, followed by two rounds of washing with phosphate-buffered saline (PBS). In total, 2 ml of cell permeabilization buffer (0.1% saponin and 0.05% sodium azide in PBS; Sigma-Aldrich) was added to the PBMCs. The PBMCs were kept on ice for 15 min in the dark, washed once with cell permeabilization buffer and resuspended in 0.2 ml of the cell permeabilization buffer. The expression of CD3ζ was detected intracellularly, using a specific phycoerythrin (PE)-conjugated monoclonal antibody (clone, 2H2D9; ca no. IM3169; dilution, 1:10; Beckman Coulter, Inc., Brea, CA, USA). An immunoglobulin (Ig)-G1 isotype control (eBiosciences, San Diego, CA, USA) was used to distinguish specific CD3ζ binding from non-specific binding during flow cytometry. The optimal amount of antibody was determined by titration and the cells were analyzed by BD FACSCanto™II flow cytometer (BD Biosciences). A total of 10,000 events were acquired and the expression levels of CD3ζ in T cells (gated on CD3ε+) were analyzed using FlowJo software, version 9.6.2 (FlowJo, LLC, Ashland, OR, USA) by the mean florescence intensity (MFI).

Freezing and thawing of PBMCs

PBMCs were isolated from the whole blood of healthy donors and HNSCC patients using Ficoll gradient centrifugation, as previously described (15,16), and washed once in complete medium, consisting of RPMI-1640, penicillin/streptomycin/L-Glutamine and 10% fetal bovine serum (Thermofisher Scientific, Inc., Ontario, Canada). Viable cell numbers were determined by a B4005 hemocytometer and trypan blue viable cell exclusion dye (VWR International, Ontario, Canada) (17). PBMCs were cryopreserved at −196°C in aliquots in complete medium, which contained a final concentration of 10% dimethyl sulfoxide, using a Mr. Frosty™ Freezing Container (Thermo Fisher Scientific, Inc., Waltham, MA, USA), according to the manufacturer's protocol. The cryopreserved PBMC was removed from liquid nitrogen, and rapidly thawed in a 37°C water bath prior to the cells being transferred to 10 ml pre-warm complete media. The cell pellet was collected and resuspended in 1 ml of fresh, warm complete media at 37°C. The viable cell count was determined, and the thawed cells were used directly for flow cytometric assays.

Internal control for flow cytometry

The present study collected PBMCs from 5 healthy donors. The PBMCs of the healthy donors were pooled and stored in small aliquots in liquid nitrogen until use. In each flow cytometric assay, one of these vials was thawed (0.5×106 cells) and used in the staining process along with the patient samples. To evaluate the intra and inter-assay variation, the same internal control was thawed and the CD3ζ expression was analyzed 3 times using flow cytometry at 3 time points in weekly intervals. The MFI of the CD3ζ chain expression was measured within an individual experiment for intra-assay variability and between multiple experiments for inter-assay variability.

Determination of cell viability and exclusion of dead cells from the analysis

To determine the cell viability and to exclude the dead cells from the analysis, Fixable Viability Dye eFluor® 780 (eBioscience, Inc., San Diego, CA, USA) was used, according to the manufacturer's protocol. The cells were washed with flow buffer (0.05% FBS, 10% sodium azide in PBS) and surface staining was performed using an anti-CD3ε monoclonal antibody. The cells were then kept in the dark at 4°C for an additional 30 minutes prior to intracellular staining for CD3ζ. Live T cells, gated from the dead cells using either the fixable viability dye or forward scatter/side scatter (FSC/SSC), were analyzed for CD3ζ chain expression.

Data normalization

CD3ζ expression (expressed as the MFI) was determined in the PBMCs of the internal control, healthy controls and patients using an individual flow cytometric assay. Data normalization was performed by transforming the MFI values of the test samples (patients and healthy donor) to a common scale using the following equation: Final relative fluorescence intensity = MFI of the test sample/MFI of the internal control.

Normality and statistical analysis

Normality and statistical analyses of the data were performed using the GraphPad Prism 5 software (GraphPad Software, Inc., La Jolla, CA, USA), termed the normality test. The D'Agostino-Pearson omnibus test was performed to investigate whether the data passed the normality test. An unpaired t-test was applied to the data that passed the normality test. The Mann-Whitney test was used on data that did not pass the normality test.

Results

Flow cytometric analysis of CD3ζ chain expression

To evaluate the specificity of anti-CD3 antibodies in flow cytometry, fluorochrome-conjugated isotype controls, consisting of the PE-mouse IgG1/κ antibody for CD3ζ detection and the FITC-IgG1/κ antibody for CD3ε detection, were used in the surface and intracellular staining of PBMC. The present study demonstrated that, when compared to the isotype controls, anti-CD3ζ and anti-CD3ε antibodies were specific (Fig. 1).

Freezing and thawing of PBMCs may induce cell death and affect the present analysis

Using a commercial fixable viability dye, the present study observed that the percentage of dead cells in the thawed samples were in the range of 2–23% (Fig. 2). To measure CD3ζ chain expression using MFI in live cells, the present study separated live cells from dead cells using the viability dye prior to the analysis of CD3ζ expression in the CD3ε+ T cells. In addition, the present study compared the differences in CD3ζ expression when the live cells were gated by either the viability dye or the FSC/SSC properties of the total cells acquired. The present study observed that there were no significant differences in the MFI determined by these two methods of live cell gating (Fig. 2).

Figure 2.

Live-cell gating by either viability dye or forward/scatter properties of the total cells acquired did not affect the MFI analyses of the CD3ζ chain expression. (A) MFI of the CD3ζ chain measured in the viable cells, gated by viability dye. In total, 0.5×106 cells of the thawed peripheral blood mononuclear cells were washed in PBS and resuspended in 200 µl of serum-free PBS. A total of 0.2 µl fixable viability dye was added to the cells, incubated for 30 min at 4°C prior to subsequent washing and surface and intracellular staining of CD3ε and CD3ζ, respectively. Using BD FACS flow cytometry 10,000 events were acquired. Based on FSC/SSC properties, lymphocyte populations were gated for subsequent discriminations of dead and viable cells by the Fixable Viability Dye eFluor® 780 (eBioscience, Inc.). Only the viable cells were gated in the MFI analyses of the CD3ζ chain in CD3ε+/CD3ζ+ cells. (B) MFI of the CD3ζ chain was determined by live-cell gating using the FSC/SSC properties of the total cells. Live lymphocyte populations were gated by their FSC/SSC properties. MFI of the CD3ζ chain in the CD3ε+/CD3ζ+ T cells were determined. (A and B) These experiments were generated from the same experiment, except the use of the viability dye in (A). (C) No significant difference in the MFI of the CD3ζ expression in the CD3ε+/CD3ζ+ cells was observed when the two methods of live-cell gating were compared. The statistical analysis was performed using Student's t-test (GraphPad Prism 5; 5 experiments were performed in total). MFI, mean fluorescence intensity; CD, cluster of differentiation; PBS, phosphate-buffered saline; FSC/SSC, foward scatter/side scatter; ns, non-significant.

Subsequently, the present study calculated the standard deviation for the inter-assay and intra-assay variations in MFI of CD3ζ expression using various aliquots of the identical PBMC sample thawed and assayed at various time points (weekly intervals for 3 consecutive weeks) for the inter-assay, and triplicate measurements of 1 aliquot of PBMC sample thawed and assayed in a single flow cytometric assay for the intra-assay. The ranges of the variations observed within an assay and between multiple assays were 407–412 and 407–477, respectively. The standard deviations for intra-assay and inter-assay were 2.16 and 28.6, respectively (Fig. 3).

Data analysis with or without normalization

Despite efforts to perform flow cytometric assays under identical conditions, the present study observed larger inter-assay variations in the MFI values of CD3ζ expression during the last 36 months. The present study noted that the MFI values of CD3ζ expression of the identical internal control varied between 378 and 1,106. It was reasoned that an internal control provided not only a quick way to examine MFI variations acquired using various equipment or the same equipment under identical photomultiplier (PMT) voltages, but also a reference value for data normalization in the inter-assays (Table II). Therefore, the present study applied a normalization procedure to the data sets of CD3ζ expression in the PBMC samples of the patients with HNSCC, which represented a low range (MFI, 378; Table III) and a high range of CD3ζ MFI (MFI, 1,106; Table IV). In addition, the present study incorporated a data set collected under an extreme MFI (MFI, 3,106; Fig. 4; Table V) to additionally test the normalization procedure. As revealed in Tables IIIV, normalization transformed the data in the low, high and extreme MFI groups into more consistent values.

Table II.

Raw and normalized data of the MFI of CD3ζ expression in CD3+ T lymphocytes.

Table II.

Raw and normalized data of the MFI of CD3ζ expression in CD3+ T lymphocytes.

Internal controlSample


Run no.Prior to normalizationFollowing normalizationPrior to normalizationFollowing normalization
1  3781.00  2810.74
211061.0022130.71
331061.00  9990.90

[i] MFI values of the same PBMC samples (internal control or a test sample) were examined and analyzed in flow cytometry at various time-points. In total, 3 representative experiments are exhibited to illustrate the wide range of variations in the MFI values observed. Using the MFI value of the internal control obtained in each run, the raw data was normalized to represent the relative alterations in the MFI. MFI, mean fluorescence intensity; CD, cluster of differentiation.

Table III.

Data normalization of the MFI of CD3ζ on CD3+ T lymphocytes in a low MFI range prior to and following data normalization.

Table III.

Data normalization of the MFI of CD3ζ on CD3+ T lymphocytes in a low MFI range prior to and following data normalization.

Prior to normalizationFollowing normalization


ValuePatientHealthy donorPatientHealthy donor
MFI1852530.480.66
2482380.650.62
2092560.550.67
2603780.681.00
2823920.741.03
2823400.740.89
2573040.670.80
2122380.560.62

[i] The MFI value for the internal control determined using flow cytometric analysis was 378. This value was used to normalize the raw MFI data of healthy donor (n=7) and patients (n=8) samples. MFI, mean fluorescence intensity; CD, cluster of differentiation.

Table IV.

Data normalization of the MFI of the CD3ζ on CD3+ T lymphocytes in a high MFI range prior to and following data normalization.

Table IV.

Data normalization of the MFI of the CD3ζ on CD3+ T lymphocytes in a high MFI range prior to and following data normalization.

Prior to normalizationFollowing normalization


ValuePatientHealthy donorPatientHealthy donor
MFI6111,2710.551.14
5151,1060.461.00
5971,3370.531.20
989   6750.890.61
601   8800.540.79
614   8200.550.74
566   5830.510.52
471   6740.420.60

[i] The MFI value for the internal control determined using flow cytometric analysis was 1,106. This value was used to normalize the raw MFI data of healthy donor (n=7) and patient (n=8) samples. MFI, mean fluorescence intensity; CD, cluster of differentiation.

Table V.

Data normalization of the MFI of the CD3ζ on CD3+ T lymphocytes in an extreme MFI range prior to and following data normalization.

Table V.

Data normalization of the MFI of the CD3ζ on CD3+ T lymphocytes in an extreme MFI range prior to and following data normalization.

Prior to normalizationFollowing normalization


ValuePatientHealthy donorPatientHealthy donor
MFI2,0962,7580.670.88
2,0192,6940.650.86
2,3562,7010.750.86
2,2763,3640.731.08
2,0923,1060.671.00
1,9722,3580.630.75
2,2152,1080.710.67
2,1132,0800.680.66

[i] The MFI value for internal control obtained using flow cytometric analysis was 3,106. This value was used to normalize the raw MFI data of healthy donor (n=7) and patient (n=8) samples. MFI, mean fluorescence intensity; CD, cluster of differentiation.

Analysis of CD3ζ expression in healthy controls and patients with HNSCC

Previous small cross-sectional studies have reported a reduced level of the CD3ζ chain in the peripheral blood lymphocytes of patients compared with healthy controls (12,13). Therefore, the present study compared the MFI of CD3ζ expression in the healthy controls and the HNSCC patients prior to and following the normalization process. The present study measured the MFI of CD3ζ expression in the CD3ε+ T lymphocytes of the healthy control individual and patient samples and normalized the data using the internal control under identical conditions in each experiment. Prior to normalization, no significant difference was observed between the MFI of CD3ζ expression in the healthy controls and the patients with HNSCC. By contrast, following normalization, CD3ζ expression was observed to be significantly decreased in the patients with HNSCC compared with the healthy controls (Fig. 5). Subsequently, the present study validated the normalization procedure in the analyses of CD3ζ expression in a longitudinal study. The present study used longitudinal PBMC samples from a single patient to illustrate the feasibility of using the normalization procedure in a longitudinal study. As revealed in Fig. 6, the use of the internal control during normalization standardized the flow cytometric CD3ζ MFI data acquired at various time points, which in turn allowed each subsequent follow-up sample to be compared to the previous sample, including the baseline data obtained prior to the patient receiving any type of clinical treatment (Fig. 6).

Discussion

The CD3ζ chain is involved in signal transduction and T cell activation (1618). Growing evidence suggests that ζ-chain degradation in cancer and other malignancies may negatively regulate T cell function (16,19); however, there has not been sufficient evidence to validate the use of ζ-chain expression as a biomarker of disease prognosis. With the wide use of flow cytometry in clinical trials, it is essential to follow gold-standard procedure to quantify biomarkers of disease. The current study developed and validated a method that used an internal control and a normalization procedure to monitor the expression levels of CD3ζ in a HNSCC patient cohort over multiple time-points using flow cytometry.

In a review that summarized a general flow cytometry equipment setup, Maecker et al suggested certain setup controls for use in flow cytometric analysis as follows: Gating controls to define positive signals and the use of biological controls, including healthy donors to compare alterations of any factor associated with alterations in the health of patients (20). The present longitudinal study consisted of an additional internal control for inter-assay variations. PBMC collected from a healthy donor was aliquoted and cryopreserved in multiple aliquots. Each aliquot was used as an internal control; the control was prepared for anti-CD3ζ staining and run alongside other test samples under specific experimental and flow cytometric conditions.

Increasing the PMT voltage is directly proportional to the strength of signals generated from a certain amount of light, and increasing or decreasing the PMT voltage directly affects the output signals and ultimately adds complexity to result analysis (21). Therefore, it is recommended to keep the PMT voltage constant for longitudinal studies. The present study observed that the inter-assay variations assayed in the short term, such as weekly intervals, was reasonably small (Fig. 3). However, despite trying to keep the PMT voltage constant in each flow cytometric analysis, the present study observed that the same internal control produced different MFI values under the same PMT voltage at various time points in a long-term longitudinal study (Fig. 4). These variations may have arisen due to various factors, including a slight alteration in PMT voltage upon each calibration, batch-to-batch variation of the fluorochrome-conjugated anti-CD3ζ antibodies and laser sensitivity. Therefore, a standardized procedure to normalize data obtained in inter-assays was required so that the MFI values were in a consistent range for statistical analyses.

Using a small sample size of healthy control individuals and patients with HNSCC, the present study compared the MFI of CD3ζ expression in the PBMCs of these two groups, with or without applying the normalization procedure. Prior to normalization, statistical analysis indicated that the distribution of data was not normal. There was no statistically significant difference between the CD3ζ expression in the PBMCs of the healthy individuals and patients with HNSCC (Fig. 5). However, following normalization, the present study observed that the distribution of the data became normal. In accordance with previous studies (12,13), the present study identified that the downregulation of the normalized MFI of the CD3ζ chain on T cells was statistically significant compared with the healthy donors (Fig. 5). Transforming the raw data, which had various ranges of MFI, into a normalized scale allowed all the variations in MFI values to become closer in range. This eliminated, not only the effect of inconsistent values generated at multiple cytometric assays, but also enabled a direct comparison of CD3ζ expression data collected at various time-points. Consequently, this method allowed the present study to observe alterations in the expression levels of the molecule of interest using flow cytometry in a longitudinal study (Fig. 6). The standardized method of measuring CD3ζ expression in frozen PBMC using flow cytometry may allow alterations in biomarkers of interest to be tracked by flow cytometry in a large cohort of patients with HNSCC in a longitudinal study.

Acknowledgements

The present study was supported by the Manitoba Medical Service Foundation and Health Science Center Foundation, University of Manitoba (Grant no. G00314660).

References

1 

Ferlay J, Shin HR, Bray F, Forman D, Mathers C and Parkin DM: Estimates of worldwide burden of cancer in 2008: GLOBOCAN 2008. Int J Cancer. 127:2893–2917. 2010. View Article : Google Scholar : PubMed/NCBI

2 

Parkin DM, Bray F, Ferlay J and Pisani P: Global cancer statistics, 2002. CA Cancer J Clin. 55:74–108. 2005. View Article : Google Scholar : PubMed/NCBI

3 

Chau CH, Rixe O, McLeod H and Figg WD: Validation of analytic methods for biomarkers used in drug development. Clin Cancer Res. 14:5967–5976. 2008. View Article : Google Scholar : PubMed/NCBI

4 

Schmalfuss F and Kolominsky-Rabas PL: Personalized medicine in screening for malignant disease: A review of methods and applications. Biomark Insights. 8:9–14. 2013. View Article : Google Scholar : PubMed/NCBI

5 

Sexton WM, Lin Y, Kryscio RJ, Dawson DR, Ebersole JL and Miller CS: Salivary biomarkers of periodontal disease in response to treatment. J Clin Periodontol. 38:434–441. 2011. View Article : Google Scholar : PubMed/NCBI

6 

Lovestone S and Thambisetty M: Blood-based biomarkers of Alzheimer's disease: Challenging but feasible. Biomark Med. 4:65–79. 2010. View Article : Google Scholar : PubMed/NCBI

7 

Latosinska A, Bhat A and Frantzi M: Clinical proteomic biomarkers: Relevant issues on study design & technical considerations in biomarker development. Clin Transl Med. 3:72014. View Article : Google Scholar : PubMed/NCBI

8 

Mahmood S, Kanwar N, Tran J, Zhang ML and Kung SK: SHP-1 phosphatase is a critical regulator in preventing natural killer cell self-killing. PLoS One. 7:e442442012. View Article : Google Scholar : PubMed/NCBI

9 

Best LM, van Veldhuyzen Zanten SJ, Bezanson GS, Haldane DJ and Malatjalian DA: Serological detection of Helicobacter pylori by a flow microsphere immunofluorescence assay. J Clin Microbiol. 30:2311–2317. 1992.PubMed/NCBI

10 

Deptala A and Mayer SP: Detection of minimal residual disease. Methods Cell Biol. 64:385–420. 2001. View Article : Google Scholar : PubMed/NCBI

11 

Dunphy CH: Contribution of flow cytometric immunophenotyping to the evaluation of tissues with suspected lymphoma? Cytometry. 42:296–306. 2000. View Article : Google Scholar : PubMed/NCBI

12 

Paietta E: Assessing minimal residual disease (MRD) in leukemia: A changing definition and concept? Bone Marrow Transplant. 29:459–465. 2002. View Article : Google Scholar : PubMed/NCBI

13 

Bagwell CB, Clark GM, Spyratos F, Chassevent A, Bendahl PO, Stål O, Killander D, Jourdan ML, Romain S, Hunsberger B and Baldetrop B: Optimizing flow cytometric DNA ploidy and S-phase fraction as independent prognostic markers for node-negative breast cancer specimens. Cytometry. 46:121–135. 2001. View Article : Google Scholar : PubMed/NCBI

14 

de Vries E, Noordzij JG, Kuijpers TW and van Dongen JJ: Flow cytometric immunophenotyping in the diagnosis and follow-up of immunodeficient children. Eur J Pediatr. 160:583–591. 2001. View Article : Google Scholar : PubMed/NCBI

15 

Kulkarni DP, Wadia PP, Pradhan TN, Pathak AK and Chiplunkar SV: Mechanisms involved in the down-regulation of TCR zeta chain in tumor versus peripheral blood of oral cancer patients. Int J Cancer. 124:1605–1613. 2009. View Article : Google Scholar : PubMed/NCBI

16 

Whiteside TL: Down-regulation of zeta-chain expression in T cells: A biomarker of prognosis in cancer? Cancer Immunol Immunother. 53:865–878. 2004. View Article : Google Scholar : PubMed/NCBI

17 

Ioannides CG and Whiteside TL: T cell recognition of human tumors: Implications for molecular immunotherapy of cancer. Clin Immunol Immunopathol. 66:91–106. 1993. View Article : Google Scholar : PubMed/NCBI

18 

Siegel AC and Louis KS: Cell viability analysis using trypan blue: Manual and automated methods. Methods Mol Biol. 740:7–12. 2011. View Article : Google Scholar : PubMed/NCBI

19 

Mizoguchi H, O'Shea JJ, Longo DL, Loeffler CM, McVicar DW and Ochoa AC: Alterations in signal transduction molecules in T lymphocytes from tumor-bearing mice. Science. 258:1795–1798. 1992. View Article : Google Scholar : PubMed/NCBI

20 

Travis J: Do tumor-altered T cells depress immune responses? Science. 258:1732–1733. 1992. View Article : Google Scholar : PubMed/NCBI

21 

Maecker HT and Trotter J: Flow cytometry controls, instrument setup, and the determination of positivity. Cytometry A. 69:1037–1042. 2006. View Article : Google Scholar : PubMed/NCBI

Related Articles

Journal Cover

March-2016
Volume 11 Issue 3

Print ISSN: 1792-1074
Online ISSN:1792-1082

Sign up for eToc alerts

Recommend to Library

Copy and paste a formatted citation
x
Spandidos Publications style
Upreti D, Pathak A and Kung SK: Development of a standardized flow cytometric method to conduct longitudinal analyses of intracellular CD3ζ expression in patients with head and neck cancer. Oncol Lett 11: 2199-2206, 2016.
APA
Upreti, D., Pathak, A., & Kung, S.K. (2016). Development of a standardized flow cytometric method to conduct longitudinal analyses of intracellular CD3ζ expression in patients with head and neck cancer. Oncology Letters, 11, 2199-2206. https://doi.org/10.3892/ol.2016.4209
MLA
Upreti, D., Pathak, A., Kung, S. K."Development of a standardized flow cytometric method to conduct longitudinal analyses of intracellular CD3ζ expression in patients with head and neck cancer". Oncology Letters 11.3 (2016): 2199-2206.
Chicago
Upreti, D., Pathak, A., Kung, S. K."Development of a standardized flow cytometric method to conduct longitudinal analyses of intracellular CD3ζ expression in patients with head and neck cancer". Oncology Letters 11, no. 3 (2016): 2199-2206. https://doi.org/10.3892/ol.2016.4209