Use of radiofrequency ablation in the treatment of hepatocellular carcinoma: Experience of ablation protocols
- Authors:
- Published online on: September 13, 2012 https://doi.org/10.3892/etm.2012.706
- Pages: 959-961
Abstract
Contents
Introduction
Differential use of needle electrodes depending on tumor diameter
Differential use of ablation methods with different needle electrodes
When expandable needle electrodes cannot be used due to tumor location
Selection of needle electrodes based on the condition of the surrounding hepatic tissue
Conclusion
Introduction
First reported by Buscarini et al in 1992 (1) and Rossi et al in 1993 (2), radiofrequency ablation (RFA) is an easy-to-operate and minimally invasive technique that provides effective local treatment. Subsequently, Shiina et al reported that, in cases with a small number (<3) of small (<3 cm in diameter) hepatocellular carcinomas (HCC), RFA was superior in terms of recurrence, and survival rates compared with the conventional HCC treatments of percutaneous ethanol injection therapy and surgical resection (3). As the rate of adverse events was similar among the three methods, they advocated the use of RFA as a first-line treatment for HCC (3). Chen et al also investigated the recurrence and survival rates of 180 patients with small HCC (<5 cm) who had been randomly assigned to either RFA treatment or surgical resection (4). They found no significant differences in these rates between the two patient groups and thus recommended RFA due to its minimal invasiveness.
RFA was introduced into our system at Toho University Medical Center Omori Hospital in 1999, and we treat nearly 200 HCC cases annually with RFA. Although individual medical facilities use their own methods of RFA, we would like here to share our experience of RFA treatment protocols.
Differential use of needle electrodes depending on tumor diameter
Expandable LeVeen (Boston Scientific Corp., Natick, MA, USA) and monopolar Cool-tip (Covidien, Boulder, CO, USA) needle electrodes have been incorporated into our system since 1999 and 2002, respectively. Depending on tumor diameter, we perform RFA using either a LeVeen needle with an array diameter of 20, 30, 35, or 40 mm or a Cool-tip needle with a 10-mm or 20-mm non-insulated tip. Regardless of tumor size, all patients receive a single RFA treatment to avoid complications and dispersal of tumor cells due to repeated needle puncture. This protocol also offers the benefit of a shorter hospital stay. Bearing in mind the average ablative diameters afforded by individual needle electrodes and the minimal ablative margin of 5 mm or larger recommended by many studies (5), we investigated whether tumor size could serve as an indication for RFA. Based on the results, we now select needle electrodes according to tumor size (Fig. 1).
Differential use of ablation methods with different needle electrodes
Cool-tip needle with a 20-mm tip
We previously used the liver of dead swine to show that the distance between the distal edge of ablation and the tip of a Cool-tip needle electrode was ≤2 mm (6). Accordingly, when no vasculature is present in the vicinity of HCC, we perform the following two-step ablation method to ensure a sufficient ablative margin from the needle tip and to avoid dispersal of tumor cells due to needle puncture.
Two-step ablation method (Fig. 2)
Insert the needle electrode into the tumor, hold the tip at the distal edge of the tumor, and turn the power on with the initial output set to 40 W.
Increase the output by 10 W at 1-min intervals until reaching 60 W. When power roll-off has occurred twice while maintaining the output at 60 W, turn the power off.
Insert the electrode a further 5 mm, repeat the above steps starting from 40 W, and end the treatment after achieving a single roll-off at 60 W.
An ablative margin of 5 mm or larger will be ensured by advancing the tip further. A needle stopper made of Duracon for percutaneous microwave coagulation therapy is used to ensure accurate needle advancement (Fig. 3) (7). We perform RFA at low power, starting from 40 W up to a maximum of 60 W, because, unlike high-power ablation methods (8), this method prevents complications and does not affect the rate of local recurrence.
Fixed ablation method. If vasculature is present in the vicinity of HCC, inserting a Cool-tip needle down to the distal edge of tumor may result in vessel perforation. In such cases we perform a different ablation method as follows:
Hold the tip of a needle electrode at the distal edge of tumor and turn the power on with the initial output set at 40 W.
Increase the output by 10 W at 1-min intervals until reaching 60 W. When power roll-off has occurred three times while maintaining the output at 60 W, end the treatment.
Cool-tip needle with a 10-mm tip
A Cool-tip needle with a 10-mm non-insulated tip is suitable for treating tumors with a diameter of <10 mm. It is particularly useful in cases with reduced hepatic functional reserve (7,9).
Hold the tip of a needle electrode at the distal edge of the tumor and turn the power on with the initial output at 20 W.
Increase the output by 10 W at 1-min intervals until reaching 30 W. When roll-off has occurred twice while maintaining the output at 30 W, turn the power off.
Insert the electrode 5 mm further, repeat the above steps starting from 20 W, and end the treatment after achieving a single roll-off at 30 W.
(However, when vasculature is present in the vicinity of HCC, perform the stationary ablation method described above with the output starting at 20 W and perform treatment while maintaining the output at 30 W).
LeVeen needle (Fig. 4)
With LeVeen needles, we perform the following steps to avoid the dispersal of tumor cells:
Hold the tip of a needle electrode at the distal edge of the tumor, partially deploy the tines, and apply the power at the output initially set for the array diameter.
Increase the output by 10 W at 1-min intervals until the power drops by 2 W, at which time maintain that output until roll-off occurs.
Depending on the diameter of tumor, deploy the tines in a stepwise fashion (in 2–4 steps) and apply the power in the same way until the tines are fully deployed.
Continue to apply 70% of the maximum output at the distal edge of tumor.
Depending on the diameter of tumor, retract the electrode by approximately 5–10 mm, perform the same stepwise ablation with full deployment, and end the treatment.
When expandable needle electrodes cannot be used due to tumor location
When performing RFA of HCC in contact with a large blood vessel (hepatic vein, hepatic portal vein) just below the diaphragm and protruding from the distal surface of the liver, even for tumor >20 mm in diameter, an expandable LeVeen needle may perforate nearby vasculature of penetrate through the liver. In the case of HCC >20 mm in diameter, our conventional method of RFA using a Cool-tip needle with a 20-mm tip does not ensure a sufficient ablative margin, and a similar problem has been reported with the 30-mm tip (10). Therefore, for HCC >20 mm in diameter, we divide a tumor equally into three segments parallel to the direction of needle puncture. We then perform two-step ablation of the segment on the right by inserting a needle electrode into the distal end of the segment. The same steps are repeated with the segment on the left. This ensures an ablation of a wide area (11) and is useful in cases where the use of a LeVeen needle is not recommended, such as those mentioned above.
Selection of needle electrodes based on the condition of the surrounding hepatic tissue
In some cases, RFA produces an unexpectedly small ablative margin. We previously reported that individual differences in ablative zone dimensions can be predicted by the depth of microbubble collapse in the liver parenchyma in the Kupffer phase of Sonazoid-enhanced ultrasonography (Fig. 5) (12). When an ablative zone is anticipated to be small because ultrasonography findings show a large area of microbubble collapse (Fig. 6), we select an electrode that produces a wider area of ablation than what we normally require to ensure a sufficient ablative margin.
Conclusion
We described the ‘tips and tricks’ of RFA treatment that we currently practice. It is our sincere hope that reporting our experience with RFA protocols will promote safe and highly effective performance of RFA treatment at all medical facilities.