Reverse transcription loop‑mediated isothermal amplification has a high performance in the detection of SARS‑CoV‑2 in saliva samples and nasal swabs from asymptomatic and symptomatic individuals

  • Authors:
    • Rodolfo Reyes-Morales
    • Patricia Segundo-Ibañez
    • César Flores-de Los Ángeles
    • David Vizcarra-Ramos
    • Damián Iñaki Ibañez-Galeana
    • Gabriela Salas-Cuevas
    • Ángel Olvera-Serrano
    • Nancy Bibiana Pérez-Silva
    • Valeria Magali Rocha-Rocha
    • Elie Girgis El-Kassi
    • Jorge Escobedo-Straffon
    • Laura Contreras-Mioni
    • Marisol Rosas-Díaz
    • Karla María Lopez-Martinez
    • Carlos Eduardo Arias-Matus
    • Elizabeth Bautista-Rodriguez
    • Manuel Nolasco-Quiroga
  • View Affiliations

  • Published online on: July 6, 2023     https://doi.org/10.3892/etm.2023.12097
  • Article Number: 398
  • Copyright: © Reyes-Morales et al. This is an open access article distributed under the terms of Creative Commons Attribution License.

Metrics: Total Views: 0 (Spandidos Publications: | PMC Statistics: )
Total PDF Downloads: 0 (Spandidos Publications: | PMC Statistics: )


Abstract

The detection of coronavirus disease 2019 cases represents a significant challenge at the epidemiological level. Limitations exist in effectively detecting asymptomatic cases, achieving good follow‑up in hospitals without the infrastructure for reverse transcription‑quantitative PCR (RT‑qPCR) or in difficult‑to‑access areas and developing methods with the need for less invasive sampling procedures. Therefore, the present study evaluated the performance of the direct reverse transcription loop‑mediated isothermal amplification (RT‑LAMP) test for detecting severe acute respiratory syndrome coronavirus 2 (SARS‑CoV‑2) in the saliva and nasal samples of asymptomatic individuals belonging to the university population. In addition, this test was also assessed for effectiveness in symptomatic individuals referred from a hospital with poor infrastructure in molecular biology and located outside the urban area. The RT‑LAMP assay was compared with the results obtained from the RT‑qPCR nasopharyngeal swab test, where the diagnosis was confirmed by lateral flow immunoassay test for rapid antigen detection. A total of 128 samples were analyzed, of which 43% were symptomatic positive individuals, 25% were asymptomatic positive individuals and 32% were SARS‑CoV2‑negative control individuals. Among positive individuals, no differences were found between the Cq values determined by RT‑qPCR. A sensitivity of 96.5% and a specificity of 97.6% was reported for the detection of SARS‑CoV‑2 in symptomatic individuals by salivary and nasal RT‑LAMP, as well as a sensitivity of 100% and a specificity of 97.6% for the detection of SARS‑CoV‑2 in asymptomatic individuals. These findings indicated that performance of the direct RT‑LAMP test using saliva and nasal samples has high sensitivity and specificity, which in turn suggest that it is a viable and reliable alternative for use in epidemiological monitoring.

Introduction

The severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), which causes coronavirus disease 2019 (COVID-19), first emerged in Wuhan, China in December 2019(1). According to the World Health Organization (WHO), the COVID-19 pandemic is the result of the rapid international spread of SARS-CoV-2. From the start of the pandemic until September 2022, ~601 million people have been infected and 6 million deaths have been recorded worldwide (database accessed on September 13, 2022; covid19.who.int/) (2). Notably, infection with SARS-CoV-2 can either remain asymptomatic or can result in severe conditions requiring hospital medical care (3,4). The transmission route of SARS-CoV-2 is typically by contact, by inhalation of respiratory droplets or by aerosols emitted by an infected individual (5). Although the number of severe cases has decreased due to vaccination, the number of cases continues to increase, although the rate of increase of cases is decreasing (6,7). Asymptomatic and symptomatic patients have similar viral loads, this is reported by Miguel et al (2022), who performed a RT-qPCR study and observed that the Ct's showed marginal differences (8). They are an important source of transmission, rendering their identification and isolation essential for containing the spread of the virus (9).

The gold standard for detecting SARS-CoV-2 is reverse transcription-quantitative PCR (RT-qPCR), a test with high sensitivity and specificity (10). However, the global increase in the number of cases revealed that, in low- and middle-income countries, there was limited accessibility to RT-PCR due to problems with sample collection, operating procedures and a lack of reagents for viral RNA extraction. The cost of the equipment, the need for trained personnel and the installation in difficult-to-access areas were some of the variables that slowed down timely detection, which consequently increased the spread of cases (11,12). As a result, other diagnostic methods have emerged, including the reverse transcription loop-mediated isothermal amplification (RT-LAMP) method. Since 2000, the advantages of RT-LAMP as a diagnostic tool in infectious diseases, such as Zika virus, dengue and malaria, have been described (13-15).

The RT-LAMP reaction is typically performed in a single step at an isothermal temperature due to the presence of the enzyme Bst or Bsm in the reaction that allows for the amplification of products together with the template RNA, reverse transcriptase and primers, without the need for specialized equipment (16). This technique uses four to six primers that bind to six regions of the specific DNA fragment, making the technique specific and sensitive. Amplified end products can then be visualized by turbidity, fluorescence, luminescence or metal-sensitive colorimetry; thus, RT-LAMP is a rapid detection method, Amaral et al (17) report that time elapsing from sample collection to delivery of results is ~30 min, depending on whether it is an RT-LAMP without RNA extraction or with RNA extraction (the latter plus the time it takes to perform RNA extraction) (17). These are more evident especially in places where the necessary equipment to perform molecular tests such as TR-qPCR is not available. Currently to perform a RT-qPCR in these places requires that the sample is taken and sent to a central laboratory for processing so it can take 2 to 5 days for diagnosis (18). RT-LAMP is considered an excellent option for rapid and continuous epidemiological follow-up in the community, in reference or specialty hospitals, in outpatients or in those living in difficult-to-access areas.

For the diagnosis of SARS-CoV-2 by RT-LAMP, a sensitivity of >80% has been reported based on saliva, nasal or nasopharyngeal samples with or without the previous step of viral RNA extraction (19,20), which holds an advantage over other tests. In Mexico, there are few diagnostic studies regarding the detection of SARS-CoV-2 with RT-LAMP. In this regard, Cisneros-Villanueva et al (21) previously reported a sensitivity and specificity of >90% from nasopharyngeal swab samples with a previous RNA extraction step. Therefore, due to the advantages offered by the RT-LAMP technique, the objective of the present study was to determine the sensitivity and specificity of the RT-LAMP test for the detection of SARS-CoV-2 in the saliva and nasal samples from symptomatic individuals derived from a hospital located far from the urban area and from asymptomatic individuals belonging to the university population. The RT-LAMP assay was compared with the results obtained by RT-qPCR of nasopharyngeal swab tests and the diagnosis was confirmed by lateral flow immunoassay (LFIA) for the rapid detection of antigens.

Materials and methods

Study design and population

In this observational and cross-sectional multi-center study, adults aged 18-70 years (mean 29.3±12.4 age) with a confirmatory diagnosis of SARS-CoV-2 determined by RT-qPCR were invited to participate, the samples were collected from January to May 2022 and samples were eliminated if there was insufficient material or if the RT-qPCR test was inconclusive. The present study had three study groups: Symptomatic patients 21 men and 35 women (mean 38.4±13.7 age), asymptomatic patients 15 men and 17 women (mean 22±5.7 age) and healthy individuals patients 19 men and 22 women (41 individuals, mean 29.7±12.4 years). Symptomatic patients were diagnosed according to the operational definition for COVID-19 issued by the Ministry of Health (México) and the WHO (22,23), which establishes that they are patients who present at least one of the following symptoms: Fever, cough or headache. In addition, they should be accompanied by at least two of the following symptoms: Dyspnea, arthralgia, myalgia, odynophagia, rhinorrhea and conjunctivitis (22). They would also have to be confirmed positive for SARS-CoV-2 by commercial RT-qPCR and antigen test (LFIA, with this test the SARS-CoV-2 antigens are recognized for antibody anti-SARS-CoV-2.),. The asymptomatic group is comprised of individuals who did not present any symptoms of COVID-19 but were tested positive for SARS-CoV-2, as determined by commercial RT-qPCR and antigen test (LFIA). The control group without COVID-19 were individuals without symptoms and who were tested negative for SARS-CoV-2, as determined by RT-qPCR and antigen test (LFIA).

Symptomatic patients were recruited from the Institute of Social Security and Services for State Workers (ISSSTE) Hospital (Huauchinango, Mexico). The asymptomatic individuals were recruited from random screening carried out in the community of the Puebla Popular University of the State (UPAEP; Puebla, Mexico). For the samples of healthy individuals were also randomly collected at UPAEP.

Sample collection and transportation

All samples including the controls were collected between January 12 and May 1, 2022. A total of five samples were taken from each participant: Two nasopharyngeal swab samples for RT-qPCR and antigen testing (LFIA), two nasal samples for RT-LAMP and antigen testing, and one saliva sample for RT-LAMP, The samples were collected when the patients presented less than 9 days of symptoms. The nasopharyngeal swab samples for RT-qPCR was placed in 15-ml tubes with 3 ml viral transport medium, whereas nasal and saliva samples were placed in 15-m tubes without any buffer. All samples were stored at a temperature of 7˚C for further processing. The samples were placed in a viral transport medium (sterile phosphate-buffered saline PBS Cat. No. 70011044 Gibco) and transported with triple packaging as established in the ‘Operative Guide for the Clinical Management of Severe Acute Respiratory Infection by COVID-19’ of the Mexican government in 2020(23).

RNA extraction for RT-qPCR

RNA was extracted from the nasopharyngeal swabs using the commercial RNeasy kit (cat. no. 74104; Qiagen GmbH) according to the manufacturer's protocol. Briefly, lysis buffer (Buffer RLT) and 10 µl proteinase K were added and sample was incubated at 55˚C for 30 min, after which 70% ethanol was added in a 1:1 volume. The sample was then transferred to the RNeasy Mini spin column, centrifuged for 15 sec at 4,000 x g at 4˚C, Subsequently 700 µl RW1 buffer (Qiagen) was added to the RNeasy column and centrifuged for 15 sec at 4,000 x g at 4˚C, For washing, the column was transferred to a new tube and 500 µl of RPE buffer (Qiagen) was added, after which it was centrifuged at 4,000 x g for 15 sec at a temperature of 4˚C, this wash was performed twice, but in the second wash it was centrifuged for 2 min. Finally the column was transferred to a 1.5 ml collection tube, 50 µl of water were added, incubated for 1 min at room temperature and finally centrifuged at 4,000 x g for 1 min. RNA quantification was performed by spectrophotometry using the Nanodrop 2000 (Nanodrop; Thermo Fisher Scientific, Inc.), to verify the integrity, an electrophoresis was performed in gel of agarose at 1.5% and SYBR® Green was used for visualization (SYBR® Green I nucleic acid gel stain; cat no. 163795-75-3 Sigma-Aldrich; Merck KGaA).

RT-qPCR for SARS-CoV-2

Samples from the ISSSTE Hospital were processed at the state laboratory of public health of Puebla, Mexico. For the detection of SARS-CoV-2, VIASURE SARS-CoV-2 Real Time PCR Detection Kit (cat. no. VS-NCO213H; Certest Biotec) was used according to the manufacturer's manual. 15 µl rehydration buffer solution was added to the PCR plates (cat. no. VS-NCO213H, the tubes contained a mixture of enzymes, primers-probes, buffer and dNTPs ), 5 µl each sample, the positive control and negative control were added and centrifuged briefly for 10 sec. PCR tubes were placed in the thermal cycler and the conditions for RT-qPCR were as follows: One cycle of 15 min at 4˚C, 1 cycle of 2 min at 95˚C, 45 cycles of 10 sec at 95˚C and 50 sec at 60˚C. before the results were uploaded onto the SISVER database (sisver.sinave.gob.mx/influenza/search_patients.php). Samples from the UPAEP were processed at the UPAEP Molecular Diagnostic Laboratory, samples were processed using a commercial kit (Logix Smart™ 2019-nCoV Kit cat. no. COVID-K-001-250-I; Co-Diagnostics), To perform RT-qPCR, 5 microliters master mix was added to the PCR tubes, 5 µl of the sample was added and centrifuged for 10 sec. The tubes were then placed in the thermal cycler (Applied Biosystems™ 7500 Fast Dx Real-Time PCR; Applied Biosystems; Thermo Fisher Scientific, Inc.) with the following conditions, 1 cycle at 45˚C for 15 min, 1 cycle at 95˚C for 2 min, and 45 cycles at 95˚C for 3 sec and 55˚C for 32 sec. In all cases, those results with a Cq value <35 were considered positive. This kit is Food and Drug Administration (US FDA) approved for the diagnosis of SARS-CoV-2 infections in the clinic, which is a RT-qPCR technique using patented CoPrimers™ technology (24).

RT-LAMP assay

For the colorimetric RT-LAMP test for detecting SARS-CoV-2, a commercial test was used, RT-LAMP SARS-CoV-2 (cat. no. BMLAMP01; Amunet) that can detect the Nucleocapsid (N) gene and open reading frames 1a (ORF1a) gene of SARS-CoV-2. A set of six primers were used for each gene tested.

For the N gene, the following primers were used: Forward outer primer F3, 5'-AGATCACATTGGCACCCG-3' and reverse outer primer B3, 5'-CCATTGCCAGCCATTCTAGC-3'; loop forward primer, 5'-GCAATGTTGTTCCTTGAGGAAGTT-3' and loop backward primer, 5'-TCGTTCCTCATCACGTAGTCGC-3' and forward inner primer, 5'-TGCTCCCTTCTGCGTAGAAGCCAATGCTGCAATCGTGCTAC-3' and backward inner primer, 5'-GGCGGCAGTCAAGCCTCTTCCCTACTGCTGCCTGGAGTT-3'.

For the ORF1a gene, the following primers were used: Forward outer primer F3, 5'-AACATGGAGGAGGTGTTGC-3' and backward outer primer B3, 5'-CAAGTAGAACTTCGTGCTG-3'; loop forward primer, 5'-GTAGCTATGTAATCATCAGA-3' and loop backward primer, 5'-TTGTCGGCCCAAATGTTAAC-3' and forward inner primer, 5'-ACTACCACCCACTTTAAGTGTAACAATGCCATGCAAGTTG-3' and backward inner primer, 5'-ATCTTGCTAAACACTGTCTTCAGAAGTTGAATGTCTTCACC-3'.

Briefly, 400 µl isotonic saline solution (0.9% PISA CS Solution) was added to the nasal sample. Subsequently, 400 µl LAMP buffer was added to the saliva and nasal swab samples and mixed by vortexing for 10 sec. A microtube was prepared, which contained 5 µl 5X Master Mix LAMP and 2.5 µl primers (primer mix for gene 1 ORF1a), 2.5 µl primers (primer mix for 1 N gene), to which 5 µl sample was added and incubated for 30 min at 65˚C in a thermoblock. Subsequently, the microtube was placed in an ice bath for 30 sec. This is a colorimetric test, so it is performed by obtaining a color change (yellow is positive) and no change is negative (pink). All trials were double-blind.

LFIA test

The LFIA assay for detecting SARS-CoV-2 was performed using the SARS CoV-2 Antigen in Saliva Rapid Test (cat. no. SKU:89355; Amunet) according to the manufacturer's instructions. Nasopharyngeal swab samples were resuspended in 10 drops of the provided running solution. Subsequently, three drops of the preparation were incorporated into the sample area of the test cartridge, before the results were interpreted after 15 min. The appearance of two lines (control and test) was considered a positive result, whereas the appearance of only the control line was a negative result.

Detection limit assessment

To determine the detection limit of the RT-LAMP assay, the NATtrol™ SARS-CoV-2 [catalog number NATSARS(COV2)-ST), pGEM-ORF1a gene (cat. no. 0810624CFHI, ZeptoMetrix LLC, ORF1a gene, USA/PHC658/2021 lineage) and pGEM-N gene (cat. no. 0810624CFHI, ZeptoMetrix LLC, N gene, USA/PHC658/2021 lineage)] was used. RT-qPCR assay was performed using the Logix Smart™ 2019-nCoV commerce kit (cat. no. COVID-K-001-250-I; Co-Diagnostics), with the conditions previously described, as a control for this test, however we start from known concentrations emitted by the supplier. The minimum detection concentration was determined by using different dilutions (108, 106, 105, 104, 103, 102, 50 and 25 copy number), with 20 replicates performed for each serial dilution. RT-LAMP kit SARS-CoV-2 (cat. no. BMLAMP01) was used to perform the RT-LAMP test. The master mix was prepared with the corresponding primers as previously described in the RT-LAMP methodology and 5 microliters of sample were added with the aforementioned concentrations.

Statistical analysis

Counts and percentages were calculated for the categorical variables. The comparison of the Cq values among the groups was carried out using an individual value plot, We perform the unpaired t-student test. The efficiency of RT-LAMP as a diagnostic test was determined by calculating its sensitivity=True Positive/(True Positive + False Negative) and specificity=True Negative/(True Negative + False Positive), considering the RT-qPCR test as the gold standard. χ2 test was used to compare the asymptomatic and symptomatic patients, and Fisher's test to compare the symptoms in female and male patients. Statistical analyses were performed using SPSS v.28 (IBM Corp.) and GraphPad Prism v. 8.0.0 for Windows (Dotmatics).

Results

Characteristics of the study population

A total of 128 samples, distributed as follows, were analyzed in the present study: 55 samples from symptomatic patients with COVID-19, 32 samples from asymptomatic patients and 41 control samples from disease-free individuals. A total of 74 (57.8%) of the participants were women. Classified by age, the majority of the participants 86 (67.1%), were aged between 18 and 39 years (Table I), 75% were between 18 and 39 years old (Table I; data not shown). The asymptomatic group did not present any symptoms or signs of the disease, whereas the three most frequent symptoms were cough, headache and odynophagia in the symptomatic group (Table I).

Table I

Distribution of patients.

Table I

Distribution of patients.

CharacteristicTotalFemaleMale P-valuea
Groupn (%)n (%)n (%)0.3236
     Positive with symptoms55 (43.0%)35 (47.2%)20 (37.0%) 
     Positive without symptoms32 (25.0%)17 (22.9%)15 (27.7%) 
     Negative control41(32.0%)22 (29.7%)19 (35.8%) 
Age    
     18-39 years86 (67.1%)45 (60.8%)41 (75.9%) 
     40-59 years36 (28.1%)24 (32.4%)12 (22.3%) 
     ≥60 years6 (4.6%)5 (6.7%)1 (1.8%) 
Symptoms    P-valueb
     Cough51 (92.7%)33 (97.0%)18 (85.7%)0.6162
     Headache40 (72.7%)26 (76.4%)14 (66.6%)0.7608
     Myalgia42 (76.3%)27 (79.4)%15 (71.4%)0.7659
     Odynophagia43 (78.1%)28 (82.3%)15 (71.4%)0.7401
     Runny nose33 (60.0)%23 (67.6%)10 (47.6%)0.2702
     Fever17 (30.9%)10 (29.4%)7 (33.3%)0.7630
     Anosmia6 (10.9%)5 (14.7%)1 (4.7%)0.3995
     Diarrhea7 (12.7%)4 (11.7%)3 (14.2%)0.6960
     Dyspnea1 (1.8%)1 (2.9%)0 (0.0%)1.0000

[i] A total of 128 patients (74 female and 54 men) were recruited to the present study, of which 55 patients had symptoms (34 women and 21 men).

[ii] aχ2 test,

[iii] bFisher's test.

Cq values of symptomatic and asymptomatic individuals

A total of 87 samples were analyzed and confirmed to be positive for SARS-CoV-2 by RT-qPCR. Analyzing the Cq values, it was revealed that all positive symptomatic patients had Cq values <35, where 9.09% of the samples had Cq values between 31 and 35, 63.63% of the samples had Cq values between 21 and 30 and 27.27% of the samples had Cq values <20. Notably, 90.09% of the samples presented with Cq values <30 (Table II).

Table II

Cq values of patients, as determined by reverse transcription-quantitative PCR.

Table II

Cq values of patients, as determined by reverse transcription-quantitative PCR.

Cq value rangeAll samples (n=87)Symptomatic (n=55)Asymptomatic (n=32) P-valuea
11-20.919 (21.8%)15 (27.2%)4 (12.5%)0.1424
21-30.962 (71.2%)35 (63.6%)27 (84.3%) 
31-356 (6.8%)5 (9.0%)1 (3.1%) 

[i] aFisher's test.

In the case of asymptomatic patients, the majority of the RT-qPCR Cq values were <30. Specifically, 3.1% of the samples had Cq values >30, 84.3% of the samples had Cq values between 31 and 30 and 12.5% of the samples had Cq values <20 (Table II). When comparing the distribution of the Cq values between the symptomatic and asymptomatic patients, reduced dispersion of data was observed in the asymptomatic patients, with the majority being concentrated around a Cq value of 25 (Fig. 1). No significant difference could be found between the Cq values of the symptomatic group and the asymptomatic group.

When analyzing all of the positive samples as a whole, it was revealed that 6.9% of the samples had Cq values >30, 71.3% had Cq values between 21 and 30 and 21.8% had Cq values <20 (Table II).

Results by type of test

A total of 128 samples were analyzed by RT-qPCR, of which 68% were positive for SARS-CoV-2 and 32% were negative. For the salivary and nasal RT-LAMP assays, 67.2% of samples were positive and 32.8% were negative, but one false positive and two false negatives were found. Regarding the LFIA test of saliva samples, 60.2% of the samples were positive and 39.8% were negative, where 10 false negatives were found (Table III).

Table III

Distribution of asymptomatic and symptomatic positive patients by type of test.

Table III

Distribution of asymptomatic and symptomatic positive patients by type of test.

TestAsymptomatic patients (n=32)Symptomatic patients (n=55)Total (n=87)
Reverse transcription-quantitative PCR32 (100.0%)55 (100.0%)87 (100.0%)
Salivary RT-LAMP32 (100.0%)53 (96.4%)85 (67.2%)
Nasal RT-LAMP32 (100.0%)53 (96.4%)85 (67.2%)
Salivary lateral flow immunoassay24 (75.0%)53 (96.4%)77 (88.5%)

[i] Data are presented as n (%). RT-LAMP, reverse transcription loop-mediated isothermal amplification.

Sensitivity and specificity of salivary and nasal RT-LAMP

The sensitivity and specificity of the salivary and nasal RT-LAMP test was determined using the following formula: Sensitivity=TP/(TP + FN); Specificity=TN/(TN + FP), TP are the true positive samples (positive samples by RT-qPCR), FN are the false negatives (Positive samples RT-qPCR but were negative RT-LAMP) and FP are false positives (Negative samples by RT-qPCR but were RT-LAMP positive). According to these data, for all samples (symptomatic and asymptomatic) a sensitivity of 97.7% and a specificity of 97.60% was obtained for salivary and nasal RT-LAMP (Table IV). When the analysis was specifically performed on the symptomatic or asymptomatic group, a sensitivity of 96.5% and a specificity of 97.6% for the detection of symptomatic individuals was found for salivary and nasal RT-LAMP, whereas a sensitivity of 100% and a specificity of 97.6% was obtained for the detection of asymptomatic individuals (Table IV).

Table IV

Sensitivity and specificity of nasal RT-LAMP test for the detection of severe acute respiratory syndrome coronavirus 2 compared with RT-quantitative PCR.

Table IV

Sensitivity and specificity of nasal RT-LAMP test for the detection of severe acute respiratory syndrome coronavirus 2 compared with RT-quantitative PCR.

CategoryAll samples (overall result)Symptomatic patientsAsymptomatic patientsNegative controls
A, Salivary RT-LAMP
     Positive85 (97.7%)53 (96.4%)32 (100.0%)1 (2.4%)
     Negative2 (2.2%)2 (3.6%)0 (0.0%)40 (97.6%)
     Total87 (100.0%)55 (100.0%)32 (100.0%)41 (100.0%)
DSe/DSpDSe=97.7DSe=96.5DSe=100DSp=97.6
B, Nasal RT-LAMP    
     Positive85 (97.7%)53 (96.4%)32 (100%)1(2.4%)
     Negative2 (2.2%)2 (3.6%)0 (0.0%)40 (97.6%)
     Total87 (100.0)55 (100.0%)32 (100.0%)41 (100.0%)
DSe/DSpDSe=97.7DSe=96.5DSe=100.0DSp=97.6
C, Lateral flow immunoassay (nasal)    
     Positive75 (86.2%)51 (92.7%)24 (75%)1 (2.4)
     Negative12 (13.7%)4 (7.2%)8 (25%)40 (97.6%)
     Total87 (100.0%)55 (100.0%)32 (100.0)41 (100.0)
DSe/DSpDSe=87.8Dse=93.2Dse=80.0Dsp=97.6

[i] Data are presented as the n (%). Dse, diagnostic sensitivity; DSp, diagnostic specificity; RT-LAMP, reverse transcription loop-mediated isothermal amplification; RT, reverse transcription.

Detection capacity assessment

To determine the detection capacity of SARS-CoV-2 by RT-LAMP assay, different concentrations of viral particles and NATtrol™ SARS-CoV-2 was used. It was observed that, when viral particles at concentrations of 108, 106, 105, 104, 103 and 102 copy number were used, the detection of NATtrol™ SARS-CoV-2, pGEM-ORF1a gene and pGEM-N gene was 100% by RT-LAMP. When viral particles at a concentration of 50 copy number were used, detection was 95% for the pGEM-ORF1a gene, 90% for the pGEM-N gene and 95% for NATtrol. Finally, when viral particles at a concentration of 25 copy number were used, detection was 75% for the pGEM-ORF1a gene, 70% for the pGEM-N gene and 85% for NATtrol (Table V).

Table V

Reverse transcription loop-mediated isothermal amplification sensitivity for the detection of each control plasmid corresponding to its respective gene.

Table V

Reverse transcription loop-mediated isothermal amplification sensitivity for the detection of each control plasmid corresponding to its respective gene.

Gene (copy number)1081061051041031025025
ORF1a gene100%100%100%100%100%100%95%75%
N gene100%100%100%100%100%100%90%70%
NATtrol™ SARS-CoV-2100%100%100%100%100%100%95%85%

[i] SARS-CoV-2, severe acute respiratory syndrome coronavirus 2; open reading frames a1 gene (ORF1a), Nucleocapsid (N).

Discussion

The present study analyzed the performance of the commercial RT-LAMP test in detecting SARS-CoV-2 in symptomatic and asymptomatic patients from nasal and saliva samples without viral RNA extraction (direct RT-LAMP). RT-LAMP was shown to have high specificity and sensitivity (>96%) in detecting SARS-CoV-2 in both symptomatic and asymptomatic patients, suggesting it to be a robust and effective test for epidemiological monitoring, detecting cases and for preventing outbreaks. The present results are in agreement with those from other studies. González-González et al (25) previously reported a sensitivity of 92.8% and a specificity of 81% for RT-LAMP in nasopharyngeal samples, whereas Schneider et al (19) reported a sensitivity of 85.9% and a specificity of 99.5% in 443 saliva samples.

Notably, in the present study, the sensitivity and specificity values were similar for both salivary and nasal RT-LAMP tests, symptomatic DSe=96.5 asymptomatic DSe=100 and DSp=97.6. However, for the nasal sample a saline solution is also added. Therefore, these findings may be attributed to the fact that the RT-LAMP tests used in the present study consisted of a set of six primers that detect two genes, N and ORF1a, of SARS-CoV-2. This was verified by Rödel et al (26), who analyzed several commercial RT-LAMP tests for detecting SARS-CoV-2, who concluded that the sensitivity and specificity tended to improve when the N and ORF1ab genes are analyzed in combination. On the other hand, Artik in 2021, also found high sensitivity and specificity of the RT-LAMP test, using the SARS-CoV-2 ORF 8 and N genes, without the need for a previous RNA extraction step (27). For this reason we decided to use an RT-LAMP test kit that analyses two genes.

RT-LAMP test in saliva and nasal swabs showed a high sensitivity (97.7%) and specificity (97.6%) compared with the RT-qPCR test in the present study. Saliva sampling presents a minimally invasive test that avoids uncomfortable nasopharyngeal sampling. Furthermore, saliva has been shown to present viral loads similar to those found in nasopharyngeal samples (28-30). Cook et al (31) recommended the salivary RT-LAMP test to monitor COVID-19 in the general population outside of health care facilities, who also revealed that the detection capacity of SARS-CoV-2 in saliva by RT-LAMP does not differ significantly from the detection in nasopharyngeal samples, even in asymptomatic patients. In addition, it has been recommended that the RT-LAMP test be performed using saliva samples within the first 9 days of disease onset (32). In the present study, all tests of symptomatic patients were performed during this interval, this being an important variable that may be associated with the high specificity and sensitivity found in the present study.

In the present study, the commercial test did not require a viral RNA extraction step, which may facilitate the logistics in the processing of samples in places with poor access to public health services and infrastructure. The present study performed screening in asymptomatic outpatients from a university and individuals from a hospital far from the urban area with little infrastructure for the detection of SARS-CoV-2. This implemented methodology, without purification of genetic material, has also been reported for RT-LAMP in other previous studies. Kidd et al (33) reported minimal differences in the sensitivity and specificity of RT-LAMP for saliva samples, compared with those from nasal and oropharyngeal swabs. Furthermore, this previous study reported that the sensitivity and specificity values of salivary RT-LAMP were 84.62 and 100, respectively, for Cq <45, compared to sensitivity of 99.1 and specificity of 100 for Cq values <25 detected by RT-qPCR of ORF1ab (33).

Fowler et al (34) performed RT-LAMP for SARS-CoV-2 with nasopharyngeal samples, reporting a sensitivity of 97% and a specificity of 99% in samples with RNA extraction, and a sensitivity of 75% and a specificity of 97% in samples without RNA extraction. This research group associates these results with the performance of using Cq values, since they reported 100% specificity and 97% sensitivity in Cq values <25 in samples without RNA extraction, likewise they report that in samples with extraction of RNA a specificity of 100% and a sensitivity of 99%.

Furthermore, Baba et al (35) reported 98% sensitivity with Cq values <30 and demonstrated that the sensitivity decreases as Cq values increase. However, this change was not drastic, since in this previous study, a sensitivity of 97% was identified at Cq <35(35). In the present study, the analyzed samples presented Cq values between 11 and 35, and we found high efficiency of the RT-LAMP test to detect patients positive for SARS-CoV-2 . This is likely because the commercial kit in the present study detects two genes instead of one, unlike the study by Fowler et al which only detects the ORF1ab gene (34).

Notably, the present study detected 100% sensitivity for salivary and nasal RT-LAMP in the detection of asymptomatic cases, which is essential to prevent the spread of the virus, especially in those with high viral loads. These results support the findings of other groups, such as Brown et al (36), who also reported 100% sensitivity using direct RT-LAMP for the detection of SARS-CoV-2 in saliva samples, confirming that it is a viable alternative technique to RT-qPCR for detection of new cases.

Previous studies have shown the efficacy of RT-LAMP concerning the gold standard RT-qPCR technique, it is also an economical technique and fast that do not require specialized personnel (16,25,27). The advantage of not requiring prior RNA extraction is that it allows minimal handling and does not require specialized personnel. Therefore, RT-LAMP could be self-applied at home, applied in the community or in hospitals lacking in facilities for detecting SARS-CoV-2. The present study screened outpatients from the academic community and patients from a hospital located in a rural area which does not have specialized equipment to perform RT-qPCR. Additionally, the present study revealed that salivary RT-LAMP has high sensitivity and specificity. The use of this type of sample represents an advantage by reducing invasiveness (32,33). In addition, pilot studies could be implemented in difficult-to-access geographical areas for the identification of vulnerable groups (20).

In conclusion, the present study revealed that RT-LAMP of SARS-CoV-2 may be considered a promising, low-cost tool that can be applied in regions that do not have specialized equipment. In addition, RT-LAMP had high sensitivity and specificity in both symptomatic and asymptomatic patients. Finally, no difference was detected between nasopharyngeal samples and saliva samples, so it is more comfortable for patients to take a saliva sample for an RT-LAMP test to detect SARS-CoV-2 virus infection in comparison to taking a nasopharyngeal swab for an RT-qPCR test.

Acknowledgements

The authors would like to thank Ms Laura Alfaro Hernández (Laboratory of Oncoimmunology and Cytomics of Childhood Cancer, Eastern Biomedical Research Center [CIBIOR], Mexico) for their technical assistance and academic support. They also thank the Council of Science and Technology of the state of Puebla (Concytep), for their support for the dissemination of this work.

Funding

Funding: No funding was received.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Authors' contributions

RRM and PSI performed the experiments, analyzed and interpreted data, and wrote the manuscript, MNQ and EBR confirm the authenticity of all the raw data. DVR, DIG, CFA, GSC, AOS, NPS, VRR, EEK, JES, LCM, KLM and CAM conducted nasal, saliva and pharyngeal sample collection and performed the routine LFIA. MRD, EBR and MNQ acquired, analyzed and interpretated the data. MRD made substantial contributions to the conception of the study. MRD, EBR and NQM designed the experiments and wrote the manuscript. All authors have read and approved the final manuscript.

Ethics approval and consent to participate

This protocol was approved by the Research Ethics Committee of the Dean of Health Sciences of the Popular Autonomous University of the State of Puebla (approval no. CONBIOETICA21CEI00620131021) and the Institute of Security and Social Services for Service Workers (approval no. 09-CEI-019-01607729). Patients provided written informed consent for participation. All procedures in studies involving human participants were performed in accordance with the ethical standards of the institutional and national research committee and the 1964 Declaration of Helsinki and its later amendments or comparable ethical standards.

Patient consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

References

1 

Guan WJ, Ni ZY, Hu Y, Liang WH, Ou CQ, He JX, Liu L, Shan H, Lei CL, Hui DSC, et al: Clinical characteristics of coronavirus disease 2019 in China. N Engl J Med. 382:1708–1720. 2020.PubMed/NCBI View Article : Google Scholar

2 

World Health Organization (WHO): Coronavirus (COVID-19) Dashboard. WHO, Geneva, 2022. http://covid19.who.int/. Accessed September 13, 2022.

3 

Wu Z and McGoogan JM: Characteristics of and important lessons from the coronavirus disease 2019 (COVID-19) outbreak in China: Summary of a report of 72314 cases from the Chinese center for disease control and prevention. JAMA. 323:1239–1242. 2020.PubMed/NCBI View Article : Google Scholar

4 

Singh S, McNab C, Olson RMK, Bristol N, Nolan C, Bergstrøm E, Bartos M, Mabuchi S, Panjabi R, Karan A, et al: How an outbreak became a pandemic: A chronological analysis of crucial junctures and international obligations in the early months of the COVID-19 pandemic. Lancet. 398:2109–2124. 2021.PubMed/NCBI View Article : Google Scholar

5 

Jayaweera M, Perera H, Gunawardana B and Manatunge J: Transmission of COVID-19 virus by droplets and aerosols: A critical review on the unresolved dichotomy. Environ Res. 188(109819)2020.PubMed/NCBI View Article : Google Scholar

6 

Abdullah F, Myers J, Basu D, Tintinger G, Ueckermann V, Mathebula M, Ramlall R, Spoor S, de Villiers T, Van der Walt Z, et al: Decreased severity of disease during the first global omicron variant covid-19 outbreak in a large hospital in tshwane, South Africa. Int J Infect Dis. 116:38–42. 2022.PubMed/NCBI View Article : Google Scholar

7 

Jassat W, Abdool Karim SS, Mudara C, Welch R, Ozougwu L, Groome MJ, Govender N, von Gottberg A, Wolter N, Wolmarans M, et al: Clinical severity of COVID-19 in patients admitted to hospital during the omicron wave in South Africa: A retrospective observational study. Lancet Glob Health. 10:e961–e969. 2022.PubMed/NCBI View Article : Google Scholar

8 

Gunatilaka AB, Marco N, Read GH, Sweeney M, Regan G, Tsang C, Abdulrahman L, Ampabathina S, Spindler A, Lu SS, et al: Viral burden and clearance in asymptomatic COVID-19 patients. Open Forum Infect Dis. 9(ofac126)2022.PubMed/NCBI View Article : Google Scholar

9 

Zou L, Ruan F, Huang M, Liang L, Huang H, Hong Z, Yu J, Kang M, Song Y, Xia J, et al: SARS-CoV-2 viral load in upper respiratory specimens of infected patients. N Engl J Med. 382:1177–1179. 2020.PubMed/NCBI View Article : Google Scholar

10 

Munne K, Bhanothu V, Bhor V, Patel V, Mahale SD and Pande S: Detection of SARS-CoV-2 infection by RT-PCR test: Factors influencing interpretation of results. Virusdisease. 32:187–189. 2021.PubMed/NCBI View Article : Google Scholar

11 

Feng W, Newbigging AM, Le C, Pang B, Peng H, Cao Y, Wu J, Abbas G, Song J, Wang DB, et al: Molecular diagnosis of COVID-19: Challenges and research needs. Anal Chem. 92:10196–10209. 2020.PubMed/NCBI View Article : Google Scholar

12 

Promlek T, Thanunchai M, Phumisantiphong U, Hansirisathit T, Phuttanu C, Dongphooyao S, Thongsopa W and Nuchnoi P: Performance of colorimetric reverse transcription loop-mediated isothermal amplification as a diagnostic tool for SARS-CoV-2 infection during the fourth wave of COVID-19 in Thailand. Int J Infect Dis. 116:133–137. 2022.PubMed/NCBI View Article : Google Scholar

13 

Silva S, Pardee K and Pena L: Loop-Mediated Isothermal Amplification (LAMP) for the Diagnosis of Zika Virus: A Review. Viruses 19, 2019.

14 

Sahni AK, Grover N, Sharma A, Khan ID and Kishore J: Reverse transcription loop-mediated isothermal amplification (RT-LAMP) for diagnosis of dengue. Med J Armed Forces India. 69:246–253. 2013.PubMed/NCBI View Article : Google Scholar

15 

Zebaze Temgoua Kemleu SG, Ngando L, Nguekeng E, Fogang B, Mafo Kapen M, Fopa SI, Biabi MF, Essangui E, Assob Nguedia JC and Ayong L: Diagnostic performance of a rapid whole blood-based RT-LAMP method for malaria diagnosis among apparently healthy blood donors and febrile neonates in Cameroon. PLoS One. 16:1–7. 2021.PubMed/NCBI View Article : Google Scholar

16 

Jamwal VL, Kumar N, Bhat R, Jamwal PS, Singh K, Dogra S, Kulkarni A, Bhadra B, Shukla MR, Saran S, et al: Optimization and validation of RT-LAMP assay for diagnosis of SARS-CoV2 including the globally dominant delta variant. Virol J. 18:1–14. 2021.PubMed/NCBI View Article : Google Scholar

17 

Amaral C, Antunes W, Moe E, Duarte AG, Lima LMP, Santos C, Gomes IL, Afonso GS, Vieira R, Teles HSS, et al: A molecular test based on RT-LAMP for rapid, sensitive and inexpensive colorimetric detection of SARS-CoV-2 in clinical samples. Sci Rep. 11:1–12. 2021.PubMed/NCBI View Article : Google Scholar

18 

Heithoff DM, Barnes LV, Mahan SP, Fox GN, Arn KE, Ettinger SJ, Bishop AM, Fitzgibbons LN, Fried JC, Low DA, et al: Assessment of a smartphone-based loop-mediated isothermal amplification assay for detection of SARS-CoV-2 and influenza viruses. JAMA Netw Open. 5(e2145669)2022.PubMed/NCBI View Article : Google Scholar

19 

Schneider FS, Molina L, Picot MC, L'Helgoualch N, Espeut J, Champigneux P, Alali M, Baptiste J, Cardeur L, Carniel C, et al: Performances of rapid and connected salivary RT-LAMP diagnostic test for SARS-CoV-2 infection in ambulatory screening. Sci Rep. 12(2843)2022.PubMed/NCBI View Article : Google Scholar

20 

da Costa VD, Santos AC, da Silva LL, Wiggers WJ, Ivantes CAP, Lima DM, Colares JKB, Dallacqua DSV, Motta-Castro ARC, Dávila AMR, et al: RT-LAMP multicenter study for SARS-CoV-2 genome molecular detection in Brazilian swab and saliva samples. Diagnostics (Basel). 13(210)2023.PubMed/NCBI View Article : Google Scholar

21 

Cisneros-villanueva M, Blancas S, Cedro-tanda A, Almaraz-rojas O, Ortiz-DR Tolentino-L, et al: Swab samples shows that it is a useful screening method for detecting, 2022.

22 

World Health Organization (WHO): Public health surveillance for COVID-19: Interim guidance. WHO, Geneva, 2020. https://apps.who.int/iris/handle/10665/333752. Accessed August 7, 2020.

23 

Secretaría de Salud: Guía operativa para el manejo clínico de la infección respiratoria aguda grave por COVID-19, 2020.

24 

Jerbi L, Azrad M and Peretz A: Evaluation of factors that affect the performance of COVID-19 molecular assays including presence of symptoms, number of detected genes and RNA extraction type. Mol Diagn Ther. 26:229–238. 2022.PubMed/NCBI View Article : Google Scholar

25 

González-González E, Lara-Mayorga IM, Rodríguez-Sánchez IP, Zhang YS, Martínez-Chapa SO, Santiago GT and Alvarez MM: Colorimetric loop-mediated isothermal amplification (LAMP) for cost-effective and quantitative detection of SARS-CoV-2: The change in color in LAMP-based assays quantitatively correlates with viral copy number. Anal Methods. 13:169–178. 2021.PubMed/NCBI View Article : Google Scholar

26 

Rödel J, Egerer R, Suleyman A, Sommer-Schmid B, Baier M, Henke A, Edel B and Löffler B: Use of the variplexTM SARS-CoV-2 RT-LAMP as a rapid molecular assay to complement RT-PCR for COVID-19 diagnosis. J Clin Virol. 132(104616)2020.PubMed/NCBI View Article : Google Scholar

27 

Artik Y, Coşğun AB, Cesur NP, Hızel N, Uyar Y, Sur H and Ayan A: Comparison of COVID-19 laboratory diagnosis by commercial kits: Effectivity of RT-PCR to the RT-LAMP. J Med Virol. 94:1998–2007. 2022.PubMed/NCBI View Article : Google Scholar

28 

Zhu Y, Wu X, Gu A, Dobelle L, Cid CA, Li J and Hoffmann MR: Membrane-based in-gel loop-mediated isothermal amplification (mgLAMP) system for SARS-CoV-2 quantification in environmental waters. Environ Sci Technol. 56:862–873. 2022.PubMed/NCBI View Article : Google Scholar

29 

Savela ES, Viloria Winnett A, Romano AE, Porter MK, Shelby N, Akana R, Ji J, Cooper MM, Schlenker NW, Reyes JA, et al: Quantitative SARS-CoV-2 viral-load curves in paired saliva samples and nasal swabs inform appropriate respiratory sampling site and analytical test sensitivity required for earliest viral detection. J Clin Microbiol. 60(e0178521)2022.PubMed/NCBI View Article : Google Scholar

30 

Tan SH, Allicock O, Armstrong-Hough M and Wyllie AL: Saliva as a gold-standard sample for SARS-CoV-2 detection. Lancet Respir Med. 9:562–564. 2021.PubMed/NCBI View Article : Google Scholar

31 

Cook BWM, Kobasa K, Tamayo M, Theriault N, Pappas DG and Theriault SS: Implementation of a rapid RT-LAMP saliva-based SARS-CoV-2 testing program in the workplace. Diagnostics. 12(474)2022.PubMed/NCBI View Article : Google Scholar

32 

Nagura-ikeda M, Imai K, Tabata S, Miyoshi K, Murahara N and Mizuno T: Clinical evaluation of self-collected saliva by quantitative. J Clin Microbiol. 58:1–9. 2020.PubMed/NCBI View Article : Google Scholar

33 

Kidd SP, Burns D, Armson B, Beggs AD, Howson ELA, Williams A, Snell G, Wise EL, Goring A, Vincent-Mistiaen Z, et al: Reverse-transcriptase loop-mediated isothermal amplification has high accuracy for detecting severe acute respiratory syndrome coronavirus 2 in saliva and nasopharyngeal/oropharyngeal swabs from asymptomatic and symptomatic individuals. J Mol Diagnostics. 24:320–336, 1-17. 2022.PubMed/NCBI View Article : Google Scholar

34 

Fowler VL, Armson B, Gonzales JL, Wise EL, Howson ELA, Vincent-Mistiaen Z, Fouch S, Maltby CJ, Grippon S, Munro S, et al: A highly effective reverse-transcription loop-mediated isothermal amplification (RT-LAMP) assay for the rapid detection of SARS-CoV-2 infection. J Infect. 82:117–125. 2021.PubMed/NCBI View Article : Google Scholar

35 

Baba MM, Bitew M, Fokam J, Lelo EA, Ahidjo A, Asmamaw K, Beloumou GA, Bulimo WD, Buratti E, Chenwi C, et al: Diagnostic performance of a colorimetric RT -LAMP for the identification of SARS-CoV-2: A multicenter prospective clinical evaluation in sub-Saharan Africa. EClinicalMedicine. 40(101101)2021.PubMed/NCBI View Article : Google Scholar

36 

Brown B, O'Hara RW, Guiver M, Davies E, Birtles A, Farooq H, Verma A, Guo H, Hayden K and Machin N: Evaluation of a novel direct RT-LAMP assay for the detection of SARS-CoV-2 from saliva samples in asymptomatic individuals. J Clin Virol Plus. 2(100074)2022.PubMed/NCBI View Article : Google Scholar

Related Articles

Journal Cover

August-2023
Volume 26 Issue 2

Print ISSN: 1792-0981
Online ISSN:1792-1015

Sign up for eToc alerts

Recommend to Library

Copy and paste a formatted citation
x
Spandidos Publications style
Reyes-Morales R, Segundo-Ibañez P, Flores-de Los Ángeles C, Vizcarra-Ramos D, Ibañez-Galeana DI, Salas-Cuevas G, Olvera-Serrano Á, Pérez-Silva NB, Rocha-Rocha VM, El-Kassi EG, El-Kassi EG, et al: Reverse transcription loop‑mediated isothermal amplification has a high performance in the detection of SARS‑CoV‑2 in saliva samples and nasal swabs from asymptomatic and symptomatic individuals. Exp Ther Med 26: 398, 2023
APA
Reyes-Morales, R., Segundo-Ibañez, P., Flores-de Los Ángeles, C., Vizcarra-Ramos, D., Ibañez-Galeana, D.I., Salas-Cuevas, G. ... Nolasco-Quiroga, M. (2023). Reverse transcription loop‑mediated isothermal amplification has a high performance in the detection of SARS‑CoV‑2 in saliva samples and nasal swabs from asymptomatic and symptomatic individuals. Experimental and Therapeutic Medicine, 26, 398. https://doi.org/10.3892/etm.2023.12097
MLA
Reyes-Morales, R., Segundo-Ibañez, P., Flores-de Los Ángeles, C., Vizcarra-Ramos, D., Ibañez-Galeana, D. I., Salas-Cuevas, G., Olvera-Serrano, Á., Pérez-Silva, N. B., Rocha-Rocha, V. M., El-Kassi, E. G., Escobedo-Straffon, J., Contreras-Mioni, L., Rosas-Díaz, M., Lopez-Martinez, K. M., Arias-Matus, C. E., Bautista-Rodriguez, E., Nolasco-Quiroga, M."Reverse transcription loop‑mediated isothermal amplification has a high performance in the detection of SARS‑CoV‑2 in saliva samples and nasal swabs from asymptomatic and symptomatic individuals". Experimental and Therapeutic Medicine 26.2 (2023): 398.
Chicago
Reyes-Morales, R., Segundo-Ibañez, P., Flores-de Los Ángeles, C., Vizcarra-Ramos, D., Ibañez-Galeana, D. I., Salas-Cuevas, G., Olvera-Serrano, Á., Pérez-Silva, N. B., Rocha-Rocha, V. M., El-Kassi, E. G., Escobedo-Straffon, J., Contreras-Mioni, L., Rosas-Díaz, M., Lopez-Martinez, K. M., Arias-Matus, C. E., Bautista-Rodriguez, E., Nolasco-Quiroga, M."Reverse transcription loop‑mediated isothermal amplification has a high performance in the detection of SARS‑CoV‑2 in saliva samples and nasal swabs from asymptomatic and symptomatic individuals". Experimental and Therapeutic Medicine 26, no. 2 (2023): 398. https://doi.org/10.3892/etm.2023.12097